Research on determination of COD with silver-free catalyst by spectrophometry
- Received Date: 23/07/2008
- Accepted Date: 13/06/2008
-
Key words:
- COD /
- XS-LP catalyst /
- precision /
- accuracy /
- correlation
Abstract: The content of COD in wastewater was determined by using the method of amended HACH which without silver but with XS-LP as catalyst. The results showed that this method was comparable to the standard method used in China in terms of accuracy and precision, were less than 10% and 3% separately, at ratio of A and B was 1∶2, gross was 12 g/L. For wastewater in nature, the linear correlation coefficient with standard method is about 0.9951 and 0.9874, accord with the requirement of the monitoring skill standard.